샘플의 상태를 실시간으로 분석할 수 있는 모니터링 라만 시스템이며, 모든 부품이 단일 챔버에 내장되어 견고한 디자인과 움직이는 부품이 없는 구조를 갖추고 있습니다.
고성능 CCD를 탑재하여 높은 해상도와 높은 감도를 제공하며, 필요 사항에 따라 Single, Dual, Multiplex Raman System, Gas Phase Raman system 을 선택할 수 있습니다.
Item | Description | Features / Specifications | |
---|---|---|---|
SF18 | Lens-based grating spectrograph | · 85 mm focal length · Standard coverage (others available upon request): - 150 to 4300 cm-1 for 488, 514, and 532nm laser - 100 to 3000 cm-1 for 785nm laser - 100 to 2000 cm-1 for 830nm laser - All preset, no moving parts |
|
Detector | TE deep cooled scientific CCD | · NIR enhanced front-illuminated sensor with no-etalonging · Peak QE > 55% at 650 nm, best for 532 and 785nm excitation · Permanent vacuum · Guaranteed TE cooling to -60°C at ambient temperature · 16-bit, 2 MHz and 100 kHz read-out: - 3.5 e- rms (typical), 6 e- rms (max) - 14 e- rms (typical), 20 e- rms (max) - 0.001 e-/p/sec (typical); 0.006 e-/p/sec (max) |
|
Laser | Single | Choose one from 488, 514, 532, 785, 830nm | 50mW (higher power available upon request) 400mW multimode frequency stabilized (single mode available upon request) |
Dual | 532/785nm | ||
Multiplex | 532/785/830nm | ||
Gas-Phase | 532nm | 100W (higher power available upon request) | |
Flow Cell Assembly | FlowC-SigE (Gas-Phase Raman) | · 316 Stainless steel with sapphire windows, 2000 psi, 200 °C · ¼” compression fittings or other fittings upon request · Integrated signal enhancer (SigE) · Multiple light path to increase sensitivity |
|
Sampling Probe | Choices of probes | · Collinear design/high throughput optics · Built-in laser line clean-up filter · Built-in deep narrow notch filter · 20 mm working distance · High collection efficiency · No interference from other lights · OD > 6: maximum rejection of Rayleigh scattering and high transmission · Wide coverage from 40 cm-1 to 4400 cm-1 |
|
Fiber connector | · Convenient SMA 905 or FC | ||
User Interface | Computer Operating System SpectraSoft | PC (Windows 10, 64bit) · Control of multiplex lasers and laser power, CCD gain and digitization, system calibration · Data processing: proprietary automatic background removal, spectrum averaging, normalization, overlay · Data analysis: peak identification, area, online monitoring |
|
Physical | Width × Depth × Height | · 300mm x 396mm x 164mm (Single, Dual, Multiplex Raman) · 300mm x 432mm x 164mm (Gas-Phase Raman) |
|
Electrical | Input Voltage | · 100 - 240 V, 50 - 60 Hz |
Background Removal(BR) 기능을 이용하여 수작업으로 제거가 불가능한 공정상의 Monitoring 등에서 자동으로 불필요한 Signal의 Background Removal을 수행할 수 있습니다.
BR 기능을 사용하여 유기 용매, 의약품, 식품성분, 고분자와 같은 대다수 샘플의 Background Signal을 신뢰성 있고 재현 가능하게 제거합니다.